Neocision 20L Chiller (-40c) - ETL Rated
Description
NEOCISION 20L Chiller (-40c) - ETL Rated
Product Name | Circulating Chiller | |||
---|---|---|---|---|
Model | NEO20L-40 | |||
Reservoir volume L) | 20L | |||
Refrigeration performance |
No-load minimum temperature (℃) |
A: ﹣ 20℃ ( Customizable: ﹣ 30℃ / C: ﹣ 40℃ | ||
Best operating temperature (℃) |
A: ﹣ 20℃ ~ 20℃ B: ﹣ 30℃ ~ 20℃ C: ﹣ 40℃ ~ 20℃ |
|||
Best operating environment (℃) |
≤30 | |||
Environmental relative humidity (%RH) | ≤65 | |||
Cooling capacity at -40 ℃ kw | 0.7 | |||
Refrigerant | R404A | |||
Electrical performance |
Cooling unit power (W) | 1500 | ||
Sensor | pt100 | |||
Controller, monitor | liquid crystal | |||
Computer interface | R485 optional | |||
safety protection | Compressor delay protection, machine leakage, over-current, over-voltage protection |
|||
Power | AC230V±10%/60HZ | |||
Pump | Power (W) | 15 | ||
Maximum flow (L/min) | 10 | |||
Maximum lift (m) | 2.5 | |||
Liquid inlet and outlet pipe diameter (mm) | φ18 | |||
Whole machine function |
Total Weight (Kg) | 105 | ||
Tank opening/depth ( mm) | φ300mm*300mm | |||
Out size (mm) | 420*580*790 | |||
Adaptation rotary evaporator (volume* number) |
10L*2 | |||
Note | The equipment can be customized to different supply voltage frequency; different flow head external circulation system and external magnetic stirring device. |
- TEMP: Displays the measured temperature value. To set the desired temperature, press the setting key until the TEMP SET value flashes. Press the setting key again to confirm the parameter.
- SET: Displays the desired temperature value set manually.
- RUN: Indicates that the chiller is operating.
- Cooling Indicator: Appears when the chiller is cooling.
- Circulation Indicator: Appears when the pump is operating.
- TIME: Displays the set running time; this value countdowns to zero and operation stops. To set a timer, press the setting key. When the TIME value flashes, enter the desired time using they arrow keys. Leave at zero if a timer is not needed.
- Silence Indicator: Appears when an alarm has been silenced.
- Alarm Indicator: It will appear when there is an alarm.
- Settings: Used for manually setting values and parameters.
- Left Arrow: Shifts to the numbers on the left to set desired value.
- Down Arrow: One push will decrease 1 digit; A long press will decrease digits faster.
- Up Arrow: One push will increase 1 digit; A long press will increase digits faster.
- Cooling Key: Press and hold key for more than 4 seconds to control start/stop the cooling function.
- Circulation Key: Press to start/stop the circulation function.
Note: After all parameters have been input, press the cooling key for four seconds to being operation.
Instructions Before Use
The unit should be used under the following conditions:
- Place on a stable, non-combustible flat surface
- Ambient temperature of 5 ~ 35 ℃
- Relative humidity ≤ 85%
- Atmospheric pressure: (86 ~ 106) KPa
- Altitude not higher than 2000 meters
- Power: 5L AC 110V±10%/60HZ, 20L: AC230V±10%/60HZ
- Keep away from any heat, corrosive gasses, strong vibration and electromagnetic sources.
- Provide a minimum of 300mm of space for proper ventilation (front and rear)
- Do not spill liquid anywhere other than the bath.
- When setting temperatures pay attention to freezing points (ethanol/ethylene glycol).
- When operating for the first time after replacing the heat transfer fluid there may be air trapped in the circulating pump, follow these instructions to purge:
- a. Disconnect the outlet line from the application NOTE: Fluid may leak from application, be ready with a bucket or plug the port. Hold the Outlet hose above the top of the bath so fluid does not drain out.
- b. Place the outlet hose into a clean collection bucket.
- c. Hold the outlet hose firmly so it does not fall out of the collection bucket
- d. Wearing safety glasses to shield any splashes, turn on the circulating function.
- e. After the liquid flow significantly increases, turn off the circulator.
- f. Reconnect the hoses to your application’s ports.
Neocision 20L Chiller (-40c) - ETL Rated
SKU: NEO20L-40
$6,870.49
You will earn 6,870.49 points for this purchase.✔ 1 Year Limited Warranty
- Want it cheaper? Call us for a wholesale or processor discount. 331-281-0154